Hardware dedicated for electrochemical measurement sometimes does not allow performing certain voltammetry experiments. Staircase and square wave voltammetry, techniques which have many advantages comparing with standard voltammetric techniques, are rarely implemented in software controlling electrochemical measuring equipment. These methods, however, can be relatively easy realized with some addition hardware and proper software. In this chapter such system and software, dedicated for cyclic, staircase and square wave voltammetry measurements, are presented. The system consists of a potentiostat and a low-cost multifunction acquisition card. Developed software allows measurement control and data acquisition. The examples of electrocatalytic gas sensor measurements using all techniques are shown.
Authors
- dr inż. Grzegorz Jasiński link open in new tab ,
- mgr inż. Anna Strzelczyk link open in new tab ,
- Łukasz Woźniak,
- Paweł Kalinowski
Additional information
- Category
- Publikacja w czasopiśmie
- Type
- artykuły w czasopismach recenzowanych i innych wydawnictwach ciągłych
- Language
- angielski
- Publication year
- 2012