In this paper a research towards understanding of mechanics of ventral hernia operated with the use of Physiomesh Open image in new window implant and SecureStrap Open image in new window staples is described. Experimental and numerical studies are conducted for that purpose. Experimental works cover uni-axial tension tests of the implant samples and of the implant-staples-tissue system. Also experiments on implant-staples-tissue models, representing operated hernia, subjected to impulse pressure loading are performed. Based on that, constitutive model of the mesh has been identified and failure load of the staples has been determined. In the experiments on the operated hernia systems subjected to pressure loading safe loading level has been determined and failure modes connected to higher pressure values have been observed. Finally, in the numerical simulations of the operated hernia model, built according to FEM rules, it has been proved that failures observed experimentally result from exceeding of the load bearing capacity of the staples considered in this study.
Authors
Additional information
- DOI
- Digital Object Identifier link open in new tab 10.1007/978-3-319-97286-2_5
- Category
- Aktywność konferencyjna
- Type
- publikacja w wydawnictwie zbiorowym recenzowanym (także w materiałach konferencyjnych)
- Language
- angielski
- Publication year
- 2018