The paper presents a preliminary meta-analysis of the sample correlation between pulse arrival time (PAT) and blood pressure (BP). The aim of the study was to verify sample correlation coefficient between PAT and BP using an affine model BP = a · P AT + b for systolic and diastolic blood pressure. The databases included in the search were the IEEE Xplore Digital Library, Springer Link and Google Scholar. Only papers from 2005 to 2017 were included into analysis. The random-effects model was considered. The resulting sample correlation coefficient was equal to -0.82 (95 % CI; -0.89, -0.72) for systolic blood pressure and -0.64 (95% CI, -0.74 -0.51) for diastolic one. Egger’s regression test showed that there was no evidence of publication bias. Obtained 95% CI intervals for sample correlation coefficients for SBP and DBP are almost separate, which may indicate different relation between PAT and BP for systolic and diastolic pressure
Authors
Additional information
- DOI
- Digital Object Identifier link open in new tab 10.1109/embc.2018.8513605
- Category
- Aktywność konferencyjna
- Type
- publikacja w wydawnictwie zbiorowym recenzowanym (także w materiałach konferencyjnych)
- Language
- angielski
- Publication year
- 2018